The G-Series High Temperature Sanitary Turbine Flowmeters are volumetric flow measurement devices. The moving fluid is used to turn a rotor, which is suspended in the flow stream. The rotating speed of the rotor is proportional to the fluid velocity or flowrate. As the blades from a spinning rotor pass by a magnetic sensor, an AC voltage pulse is generated and transmitted to the readout instrument. Each pulse is equal to a given volume of liquid therefore “x” number of pulses are equal to a gallon, litre, pound, barrel, etc.
Select a Magnetic Pick-up from The Table
Selection Guide for 1 in. 1-1/2., 2 in. Turbine Meters |
Magnetic Pickup type | Part No. | Application | Cable Type |
Wire Lead Standard | 81003000 | Meter mount electronics. Temperature range: -100°F to +250°F. Sine wave output. | None |
| | | |
Herm/High Temp. | 81002000 | Remote mount electronics. Temperature range: -450°F to +258°F. Sine wave output. | T |
High Temp. Standard | 81007000 | Remote mount electronics. Temperature range: -450°F to +800°F. Sine wave output. | T |
Active Digital | 81004000 | Remote mount electronics. Temperature range: -40°F to +248°F. Square wave, open collector output, 5-32 VDC required. Signal output: 0-10V, NPN | D |