The new OM-CP-RHTEMP1000IS-A is an intrinsicallysafe temperature and humidity data logger certified forClass I, Division 1, Groups A, B, C, and D and nonincendivefor Class I, Division 2 Groups A, B, C, Denvironments. This certification makes the device idealfor EtO sterilization, environmental studies, medical andpharmaceutical and numerous other hostile environmentapplications.
One of the enhancements to the OM-CP-RHTEMP1000IS-Ais the stainless steel enclosure with engraved serialnumber and device name. The device interfaces withthe OM-CP-IFC400 docking station or theOM-CP-IFC406 multiplexer data logger interface(sold separately) and can be placed directly intothe dock for full communications. No more removalof end-caps to download data. The small size of theOM-CP-RHTEMP1000IS-A allows it to be placed inremote or hard to reach locations. The device nowfeatures a battery life indicator allowing users to easilyidentify low voltage and change batteries when required.This feature maximizes the use of your battery life and isa cost saving efficiency.
The OM-CP-RHTEMP1000IS-A utilizes the latestsoftware. This allows for simple starting, stopping anddownloading of collected data. The devices are ableto store up to 16,350 time and date stamped humidityand temperature readings in non-volatile solid statememory. Once the readings have been downloaded tothe software, it can be viewed in graphic, tabular, andsummary form for easy analysis, as well as the potentialto be exported into Excel® for further calculations.
Specifications
Temperature
Temperature Sensor: Resistance Temperature Detector (RTD)
Temperature Range: -40 to 80°C (-40 to 176°F)
Temperature Resolution: 0.01°C
Temperature Response Time in Air:
t63 - 8 minutes to 63% of change
t90 - 68 minutes to 90% of change
Calibrated Accuracy: ±0.5°C (0 to 50°C)
Humidity
Humidity Sensor: Capacitive polymer
Humidity Range: 0 to 100% RH (non-condensing)
Humidity Resolution: 0.1% RH
Humidity Response Time in Air:
t63 - 30 seconds to 63% of change
t90 - 2 minutes to 90% of change
Calibrated Accuracy: ±3.0% RH maximum
General
Start Modes: Software programmable immediate start or delaystart up to 24 months in advance
Real Time Recording: May be used with PC to monitor andrecord data in real time
Memory: 16,350 readings per channel
Password Protection: Yes
Reading Rate: 1 reading every second up to 1 reading every24 hours
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within deviceMemory Wrap: Yes
Data Format: Date and time stamped °C, °F, K, °R ; %RH,mg/mL, dew point
Battery Life: 2 years typical at 25°C (15 minute reading rate)
Battery Type: 3.6V lithium battery included, user replaceable
Time Accuracy: ±1 minute/month at 25°C
Computer Interface: OM-CP-IFC400 USB docking station orOM-CP-IFC406 multiplexer interface required; 125,000 baud
Software: Windows® XP SP3/Vista/7 and 8 (32-bit and 64-bit)
Operating Environment: -40 to 80°C (-40 to 176°F),0 to 95% RH (non-condensing)
Dimensions: 42 L x 24.6 mm dia (1.7 x 0.97")
Weight: 65 g (2.3 oz)
Material: 316 Stainless Steel