包装
  • 559 mm (22") Flexible Probe
  • Operating Temperature Range:
    -40 to 140°C (-40 to 284°F)
  • Optional Thermal Shield
    for Operation up to 250°C (482°F)
  • Submersible NEMA 4X (IP68)
  • Programmable Start Time
  • User Calibration Through Software
  • Real-Time Operation
  • Autoclave Verification
  • Pharmaceutical
  • Implement HACCP Programs
  • Food Preparation and Processing
  • Ready to Use Right Out-of-the-Box
  • Free NIST Certification

  • The OM-CP-HITEMP140-PT Seriesare submersible, temperature dataloggers that can operate up to 140°C(284°F) and have an accuracy of±0.1°C (0.18°F). These devicesfeature a 559 mm (22") flexiblestainless steel probe for measuringextended temperatures up to 260°C(500°F). The probe is durable andcan be spiraled, bent or angled inany direction, making it easy to logtemperatures within bottles, vials orother hard to reach places.

    The OM-CP-HITEMP140-PTrecords date and time stampedreadings, and has non-volatilesolid state memory that will retaindata even if the battery becomesdischarged. Data retrieval is simple.Plug the device into an availableUSB port and our easy-to-usesoftware does the rest. The softwareconverts a PC into a real-time stripchart recorder.

    Using the software, starting,stopping and downloading theOM-CP-HITEMP140-PT is simpleand easy. Graphical, tabular andsummary data is provided foranalysis and data can be viewed in°C, °F, K or °R. The data can also beautomatically exported to Excel® forfurther calculations.

    The OM-CP-HITEMP140-TSKis a kit that includes aOM-CP-HITEMP140-PT data loggerhoused in a thermal shield.The combined features ofthe ±0.1°C accuracy of theOM-CP-HITEMP140-PT and theproperties of the durable thermalshield allow the device to be usedfor a wide range of validationapplications.Using the software, the datalogger is fast and easy to setup.Remove the thermal shield andplace the OM-CP-HITEMP140-PTinto the OM-CP-IFC400 dockingstation (sold separately). Using thesoftware, an immediate or delaystart can be chosen, as well as thereading rate. Select Start to programthe settings and start the datalogger.

    Place the thermal shield around theOM-CP-HITEMP140-PT and screwit back together. The device is readyto be deployed.The OM-CP-HITEMP140-TSK canbe completely submerged and isbuilt for applications that requireextreme temperature monitoring.These two devices can be startedand stopped directly from acomputer and their compact designallows them to fit almost anywhere.

    Specifications
    OM-CP-HITEMP140-PT-TSK
    (Thermal Shield)
    Operating Environment:

    -200 to 250ºC (-328 to 482ºF)(time limited) 0 to 100% RH
    Enclosure Material: PTFE
    Dimensions: 110 H x 51 mm dia. (4.3 x 2.0")
    Weight: 274 g (9.7 oz) not including data logger

    OM-CP-HITEMP140-PT
    (Without Thermal Shield)
    Temperature Sensor:
    100Ω Platinum RTD
    Probe Measurement Range:-200 to 260ºC (-328 to 500ºF)
    Temperature Resolution:0.01ºC (0.02ºF)
    Calibrated Accuracy:±0.1ºC (±0.18ºF); 20 to 140ºC (68 to 284ºF)
    Start Modes: Softwareprogrammable immediate startor delay start
    up to eighteen months in advance.
    Stop Modes: Manual throughsoftware timed (specific data and time)
    Real Time Recording: May beused with PC to monitor and record data in real time
    Password Protection: An optionalpassword may be programmedinto the device
    to restrict access toconfiguration options, Data may beread out with the password,
    Memory: 32,700 readings
    Wrap Around: Yes
    Reading Rate: One second up to once every 24 hours
    Battery Type: 3.6V hightemperature lithium battery(included); user replaceable
    Battery Life: 1 year typical[1 minute reading rate at 25ºC (77ºF)]
    Calibration: Digital calibration through software
    Calibration Date: Automaticallyrecorded within device
    Data Format: Date and timestamped ºC, ºF, K, ºR
    Time Accuracy: ±1 minute/monthat 20 to 30ºC (68 to 86ºF)
    Computer Interface: OM-CP-IFC400 docking stationrequired; 125,000 baud
    Software: Windows® XP SP3/Vista/7 and 8 (32- and 64-bit)
    Operating Environment:-40 to 140ºC (-40 to 284°F),0 to 100% RH
    Weight: 120 g (4.2 oz)
    Material: 316 Stainless Steel
    Dimensions:
      Body:
    48 x 25 mm dia. (1.9 x 0.97")
      Probe Tip (OM-CP-HITEMP140-PT-1): 3.2 OD x 42 mm L (0.125 x 1.5")
      Probe Tip (OM-CP-HITEMP140-PT-5): 3.2 OD x 121 mm L (0.125 x 4.8") with 4.8 OD x 25 mm L handle (0.188 x 1")
      Flexible Probe Portion: 559 x 1.6 mm OD dia.(22 x 0.062")
















暂无内容提供
暂无内容提供

联系方式

中国自动化网安全提示,在购买供应商商品时,请一定要先审核好企业信息是否真实,谨防网上诈骗。
公司名称:Omega工业测量(中国)
联系人: 王先生
联系电话:02161973664
手机号:
传 真:
电子信箱:kevin.wang@cn.omega.com
公司网站:http://cn.omega.com
公司地址:上海市宜山路1698号102单元
留言反馈
  • 评价:

  • 关于:

  • 联系人:

  • 联系电话:

  • 联系邮箱:

  • 需求意向:

  • 验证码:

    看不清楚?

  • 在线咨询
X
下载企业APP

成为企业会员免费生成APP!