OM-USB-1608G Series

OM-USB-1608G Series

分  类:其它

品  牌:欧米茄

厂  商:Omega工业测量(中国)

型  号:OM-USB-1608G Series

相关行业:

发布时间:2019-12-19 18:00 浏览人数:

最新产品

产品详情
参数
包装
  • 16-Bit High-Speed USB Devices
  • Acquisition Rates Ranging From 250 kS/s to 500 kS/s
  • 8 Differential (DIFF) or 16 Single-Ended (SE) Analog Inputs (Software Selectable)
  • Up to 2 Analog Outputs
  • 8 Digital I/O Lines
  • Two 32-Bit Counter Input Channels
  • One Timer Output Channel
  • TracerDAQ® Software Included for Acquiring and Displaying Data and Generating Signals
  • Universal Library Includes Support for Visual Studio® and Visual Studio.NET, Including Examples for Visual C++®, Visual C#®, Visual Basic®, and Visual Basic® .NET
  • Comprehensive Drivers for DASYLab® and NI LabVIEW™
  • InstaCal Software Utility for Installing, Calibrating, and Testing
  • Supported Operating Systems: Windows® XPSP2 (32-bit only) and VISTA/7/8 (32-bit and 64-bit)

  • The OM-USB-1608G Series devices are low-cost, high-speed, analogand digital I/O USB devices. All of these devices offer up to eightdifferential (DIFF) or 16 single-ended (SE) analog inputs, up to eight digitalI/O channels, two counter inputs, and one timer output.The OM-USB-1608GX-2AO offers two, 16-bit analog output channelswith DAC rates up to 500 kS/s. Everything you need to beginacquiring, viewing, and storing data is included with eachOM-USB-1608G Series device, including comprehensivesoftware support.

    Analog Input
    OM-USB-1608G Series devices provide 16-bit analog inputs thatare software-selectable as 16 SE or eight DIFF inputs. These devices alsosupport input ranges of ±10V, ±5V, ±2V, and ±1V that are software-selectableper channel

    Analog Output
    (OM-USB-1608GX-2AO only) The OM-USB-1608GX-2AO has two16-bit analog outputs. Both outputs can be updated at a rate of 250 kS/sper channel; one output can be updated at a rate of 500 kS/s. Theoutput range is fixed at ±10V. The outputs default to 0V when the hostPC is shut down or suspended, or when a reset command is issued tothe device.

    Trigger Input
    OM-USB-1608G Series devices have an external digital trigger input. Thetrigger mode is software-selectable for edge- or level sensitive mode.You can configure edge-sensitive mode for either rising or falling edge.In level sensitive mode, you can configure for either high or low level.The default setting at power up is edge-sensitive, rising edge.

    Retrigger Mode:
    In retrigger mode, you can set up repetitive analoginput trigger events. The trigger is automatically re-armed after it isactivated. Use software to set the A/D trigger count (the number of samplesyou want per trigger).

    Digital I/O
    Eight bidirectional digital I/O connections are included withOM-USB-1608G Series devices. Each digital channel is individuallyconfigurable for input or output. The digital I/O terminals can detect thestate of any TTL-level input. You can configure for pull-up (+5V) or pulldown(0V) through a jumper.

    Counter Input
    Each OM-USB-1608G Series device includes two 32-bit event countersfor counting TTL pulses. The counters accept frequency inputs ofup to 20 MHz.

    Timer Output
    OM-USB-1608G Series devices have a PWM timer output thatcan generate a pulse output with a programmable frequency in therange of 0.0149 Hz to 32 MHz. The timer output parameters aresoftware-selectable.

    External Clock I/O
    OM-USB-1608G Series devices provide one external clock inputand one external clock output for the analog inputs. TheOM-USB-1608GX-2AO also has one external clock input andone external clock output for the analog outputs.

    Software
    The OM-USB-1608G modules ship with an impressive array of software,including the new TracerDAQ®, a fullfeatured, out-of-the-box data logging,viewing, and analysis application. Driver support and detailedexample programs are included for Universal Library programminglibraries for Microsoft® Visual Studio® programming languages, and otherlanguages, including DASYLab®, and ULx for NI LabVIEW® (comprehensivelibrary of VIs and example programs compatible with 32-bit and 64-bitLabVIEW v8.5 through 2012) and InstaCalTM installation, calibrationand test utility-powerful solutions for programmers and nonprogrammersalike. These modules operate under Microsoft Windows® XP (32-bit only) and VISTA/7/8 (32-bit and 64-bit) operating systems.

    The OM-USB-1608G data acquisition modules are suppliedwith TracerDAQ software which is a collection of four virtual instrumentapplications used to graphically display and store input data andgenerate output signals:
    • Strip Chart—Log and graph valuesacquire from analog inputs, digitalinputs, temperature inputs and  counter inputs
    • Oscilloscope—Display valuesacquired from analog inputs
    • Function Generator—Generatewaveforms for analog outputs
    • Rate Generator—Generatewaveforms for counter outputs

    TracerDAQ PRO is an enhancedversion of TracerDAQ and is available as a purchased upgrade
    (SWD-TRACERDAQ-PRO). A comparisonof some of the features included inTracerDAQ vs TracerDAQ PRO isshown on the PDF.

    Specifications
    Analog Input
    A/D Converter Type: Successive approximation
    ADC Resolution: 16-bits
    Number of Channels: 8 DIFF, 16 SE; software-selectable
    Input Voltage Range: ±10V, ±5V,±2V, ±1V; software-selectable per channel
    Absolute Maximum Input Voltage
      CHx Relative to AGND: ±25Vmaximum (power on); ±15V maximum (power off)
    Input Impedance: 1 GΩ (power on); 820 Ω (power off)
    Input Bias Current: ±10 nA
    Input Bandwidth
      All Input Ranges, Small Signal (-3 dB)

      OM-USB-1608G: 750 kHz
      OM-USB-1608GX and OM-USB-1608GX-2AO: 870 kHz
    Input Capacitance: 60 pf
    Maximum Working Voltage (Signal + Common Mode):±10.2 V maximum relative to AGND
    Common Mode Rejection Ratio fIN = 60 Hz, All Input Ranges: 86 dB
    Crosstalk
    Adjacent Differential Mode Channels, DC to 100 kHz: -75 dB
    Input Coupling: DC
    Sampling Rate (Software-Selectable)
       OM-USB-1608G: 0.0149 Hz to 250 kHz
       OM-USB-1608GX and OM-USB-1608GX-2AO: 0.0149 Hz to 500 kHz
    Trigger Source: TRIG (refer to External Trigger section)
    Sample Clock Source: Internal A/D clock or external A/D clock (AICKI terminal)
    Burst Mode: Software-selectable using the internal A/D clock; alwaysenabled
    when using the external clock (AICKI terminal)
      OM-USB-1608G: 4 µs
      OM-USB-1608GX and OM-USB-1608GX-2AO: 2 µs
    Throughput
      Software Paced: 33 to 4000 S/s typical, system dependent
      Hardware Paced
      OM-USB-1608G: 250 kS/s maximum
      OM-USB-1608GX and OM-USB-1608GX-2AO: 500 kS/s maximum
    Channel Gain Queue: Up to 16 elements; software-selectablerange for each channel
    Warm-up Time: 15 minutes minimum

    Analog Output
    (OM-USB-1608GX-2AO only)

    Number of Channels: 2 (leave unused AOUTx output channelsdisconnected)
    Resolution: 16-bits
    Output Range: ±10V (calibrated)
    Output Transient
    Host computer is reset, powered on, suspended, or a reset command is issued to the device.
      Duration: 500 µs
      Amplitude: 2V pk-pk
    Powered Off
      Duration: 10 ms
      Amplitude: 7V peak
    Differential Non-Linearity: ±0.25 LSB typ; ±1 LSB maximum
    Output Current
      AOUTx: ±3.5 mA maximum
    Output Short-Circuit Protection
      AOUTx Connected to AGND: Unlimited duration
    Output Coupling: DC
    Power On and Reset State
      DACs Cleared to Zero-Scale:
      0V, ±50 mV (AOUTx defaults to 0V whenever the host computeris reset, powered on, suspended,
      or a reset commandis issued to the device)
    Output Noise: 30 µVrms
    Sample Clock Source: Internal D/Aclock or external D/A clock (AOCKI terminal)
    Output Update Rate: 500 kHz/number of channels in scan
    Settling Time
      To Rated Accuracy, 10V Step: 40 µs
    Slew Rate: 9 V/µs
    Throughput
      Software Paced: 33 S/s to 4000 S/s typical, system-dependent
      Hardware Paced: 500 kS/s maximum, system-dependent

    Analog Input/Output Calibration
    Recommended Warm-Up Time: 15 minutes minimum
    Calibration Method: Self-calibration (firmware)
    Calibration Interval: 1 year (factory calibration)
    AI Calibration Reference
    +5V, ±2.5 mV maximum (actual measured values stored inEEPROM)
      Tempco: 5 ppm/°C maximum
      Long Term Stability: 15 ppm/1000 hours
    AOUTx Calibration Procedure
    (OM-USB-1608GX-2AO)
    The analog output terminals are internally routed to the analog inputcircuit. For best calibration results, disconnect any AOUTx connectionsat the terminal block prior to performing AOUT calibration.

    Digital Input/Output
    Digital Type: CMOS
    Number of I/O: 8
    Configuration: Each bit may be configured as input (power ondefault) or output
    Pull-Up Configuration: The port has 47 kΩ resistors configurable aspull-ups or pull-downs (default)
    via internal jumper (W1)
    Digital I/O Transfer Rate (System-Paced):
    33 to 8000 port reads/writes or single bit reads/writesper second typ, system dependent
    Input High Voltage: 2.0V minimum, 5.5V absolute maximum
    Input Low Voltage: 0.8V maximum, -0.5V absolute minimum,0V recommended minimum
    Output High Voltage: 4.4V minimum (IOH = -50 µA), 3.76V minimum (IOH = -2.5 mA)
    Output Low Voltage: 0.1V maximum (IOL = 50 µA),0.44V maximum (IOL = 2.5 mA)
    Output Current: ±2.5 mA maximum

    External Trigger
    Trigger Source: TRIG input
    Trigger Mode: Software configurable for edge or levelsensitive, rising or falling edge, high or low level. Power on defaultis edge sensitive, rising edge.
    Trigger Latency: 1 µs + 1 clock cycle maximum
    Trigger Pulse Width: 100 ns minimum
    Input Type: Schmitt trigger, 33 Ω series resistor and 49.9 kΩpull-down to ground
    Schmitt Trigger Hysteresis: 0.4V to 1.2V
    Input High Voltage: 2.2V minimum, 5.5V absolute maximum
    Input Low Voltage: 1.5V maximum, -0.5V absolute minimum, 0Vrecommended minimum

    External Clock Input/Output
    Terminal Names
      OM-USB-1608G and OM-USB-1608GX: AICKI, AICKO,
      (OM-USB-1608GX-2AO): AOCKI, AOCKO
    Terminal Types
      AxCKI: Input, active on rising edge
      AxCKO: Output, power on default is 0V, active on rising edge
    Terminal Descriptions
      AxCKI: Receives sampling clock from external source
      AxCKO: Outputs the internal sampling clock (D/A or A/D clock) or the pulse generated from AxCKI   when in external clock mode.
    Input Clock Rate
      OM-USB-1608G: 250 kHz maximum
      OM-USB-1608GX and OM-USB-1608GX-2AO: 500 kHz maximum
    Clock Pulse Width
      AxCKI: 400 ns minimum
      AxCKO: 400 ns minimum
    Input Type: Schmitt trigger, 33 Ω series resistor, 47 kΩ pull-down to ground
    Schmitt Trigger Hysteresis: 0.4V to 1.2V
    Input High Voltage: 2.2V minimum, 5.5V absolute maximum
    Input Low Voltage: 1.5V maximum, -0.5V absolute minimum, 0Vrecommended minimum
    Output High Voltage: 4.4V minimum (IOH = -50 µA),3.76V minimum (IOH = -2.5 mA)
    Output Low Voltage: 0.1V maximum (IOL = 50 µA),0.44V maximum (IOL = 2.5 mA)
    Output Current: ±2.5 mA maximum

    Counter
    Terminal Names: CTR0, CTR1
    Number of Channels: 2 channels
    Resolution: 32-bit
    Counter Type: Event counter
    Input Type: Schmitt trigger, 33 Ωseries resistor, 47 kΩ pull-down to ground
    Input Source: CTR0 (terminal 52) CTR1 (terminal 51)
    Counter Read/Writes Rates (Software-Paced):
    33 to 8000 reads/writes per second typical,system dependent
    Input High Voltage: 2.2V minimum, 5.5V maximum
    Input Low Voltage: 1.5V maximum, -0.5V minimum
    Schmitt Trigger Hysteresis: 0.4V minimum, 1.2V maximum
    Input Frequency: 20 MHz, maximum
    High Pulse Width: 25 ns, minimum
    Low Pulse Width: 25 ns, minimum

    Timer Output
    Timer Terminal Name: TMR
    Timer Type: PWM output with count, period, delay, and pulsewidth registers
    Output Value: Default state is idle low with pulses high, software selectable output invert
    Internal Clock Frequency: 64 MHz
    Register Widths: 32-bit
    High Pulse Width: 15.625 ns minimum
    Low Pulse Width: 15.625 ns minimum
    Output High Voltage: 4.4V minimum (IOH = -50 µA), 3.76V minimum (IOH = -2.5 mA)
    Output Low Voltage: 0.1V maximum (IOL = 50 µA), 0.44Vmaximum (IOL = 2.5 mA)
    Output Current: ±2.5 mA maximum

    Memory
    Data FIFO: 4 kS analog input/2 kS analog output
    Non-Volatile Memory: 32 KB (28 KB firmware storage, 4 KBcalibration/user data)

    Power
    Supply CurrentThis is the total quiescent current requirement for the device thatincludes up to 10 mA for the Status LED. This does not include anypotential loading of the digital I/O bits, +5V terminal, or the AOUTxoutputs (OM-USB-1608GX-2AO only)

    Quiescent Current
    OM-USB-1608G and OM-USB-1608GX: 230 mA
    OM-USB-1608GX-2AO: 260 mA
    +5V User Output Voltage Range (Available at Terminal 42):
    4.9V minimum to 5.1V maximum
    +5V User Output Current(Available at Terminal 42):
    10 mA maximum

    Environmental
    Operating Temperature Range:0 to 55°C (32 to 131°F) maximum
    Storage Temperature Range:-40 to 85°C (-40 to 185°F) maximum
    Humidity: 0 to 90% RH non-condensing max

    Mechanical
    Dimensions:
    127 L × 89.9 W × 35.6 H mm (5.00 × 3.53 × 1.40")
    Weight: 160 g (0.35 lb)
    USB Cable: 3 m (9.84") maximum



























































































































































































































暂无内容提供
暂无内容提供

联系方式

中国自动化网安全提示,在购买供应商商品时,请一定要先审核好企业信息是否真实,谨防网上诈骗。
公司名称:Omega工业测量(中国)
联系人: 王先生
联系电话:02161973664
手机号:
传 真:
电子信箱:kevin.wang@cn.omega.com
公司网站:http://cn.omega.com
公司地址:上海市宜山路1698号102单元
留言反馈
  • 评价:

  • 关于:

  • 联系人:

  • 联系电话:

  • 联系邮箱:

  • 需求意向:

  • 验证码:

    看不清楚?

  • 在线咨询
X
下载企业APP

成为企业会员免费生成APP!