FDT-40 Series

FDT-40 Series

分  类:其它

品  牌:欧米茄

厂  商:Omega工业测量(中国)

型  号:FDT-40 Series

相关行业:

发布时间:2019-12-19 17:34 浏览人数:

最新产品

产品详情
参数
包装
  • Bi-Directional Flow Measurement
  • Rugged, Aluminum Enclosure
  • Rate and Total Backlit Display
  • 4 to 20 mA and Dual Alarm Outputs
  • USB Programming Port
  • RS485 MODBUS® Network Connection

  • FDT-40 Series ultrasonic flow andenergy meters clamp on to theoutside of pipes and do not comein contact with the internal liquid.The technology has inherentadvantages over alternate devicesincluding: low-cost installation, nopressure head loss, no moving partsto maintain or replace, no fluidcompatibility issue, and a large,bi-directional measuring range thatensures reliable readings even atvery low and high flow rates.FDT-40 Series is available in avariety of configurations thatpermit the user to select a meter withfeatures suitable to meet particularapplication requirements.

    Specifications
    System
    Liquid Types: Most clean liquids or liquids containing small amounts of suspended solids or gas bubbles
    Velocity Range: Bi-directional to greater than 40 FPS (12 MPS)
    Flow Accuracy:
        Larger Than 50 mm (2"): ±1% of reading or ±0.01 FPS (0.003 MPS), whichever is greater
        25 to 50 mm (1 to 2"): ±1% of reading or ±0.04 FPS (0.012 MPS), whichever is greater
    ¾" (19 mm) and Smaller: ±1% of full scale
    Sensitivity Flow: 0.001 FPS (0.0003 MPS)
    Repeatability: 0.5% of reading
    Transmitter
    Power Requirements (AC): 95 to 264 Vac 47 to 63 Hz @ 17 VA maximum standard, 20 to 28 Vac 47 to 63 Hz @ 0.35 A maximum (“-24VAC” models), 10 to 28 Vdc @ 5 W maximum (“-VDC” models)
    Protection: Auto resettable fuse, reverse polarity and transient suppression
    Display: Two line LCD, LED backlit; top row 18 mm (0.7") height, 7-segment; bottom row 9 mm (0.35") height, 14-segment
    Icons: RUN, PROGRAM, RELAY1, RELAY2
    Flow Rate Indication: 8-digit positive, 7-digit negative maximum; auto decimal, lead zero blanking
    Flow Totalizer: 8-digit positive, 7-digit negative maximum (re-set via keypad press, software, network command or momentary contact closure
    Enclosure NEMA 4 (IP65) Construction: Powder-coated aluminum, polycarbonate, stainless steel, polyurethane, nickel-plated steel mounting brackets
    Size (Electronic Enclosure Only): 152 W x 112 H x 56 mm D (6.0 x 4.4 x 2.2")
    Conduit Holes: (2) ½ NPT female (1) ¾ NPT female
    Temperature: -40 to 85°C (-40 to 185°F)
    Configuration: Via optional keypad or PC running software (Note: not all configuration parameters are available from the keypad – i.e. flow and temperature calibration and advanced filter settings)
    Engineering Units Flow Meter: Feet, gallons, cubic feet, million gallons, barrels (liquid and oil), acre-feet, lbs, meters, cubic meters, liters, million liters, kg
    Inputs/Outputs USB 2.0: For connection of a PC running configuration utility
        RS485: MODBUS RTU command set
        4 to 20 mA: 12-bit, internal power, can span negative to positive flow/ energy rates
        0 to 1000 Hz: Open-collector, 12-bit, can span negative to positive rates; square-wave or turbine meter simulation outputs
    Two Alarm Outputs: Open-collector, configure as rate alarm, signal strength alarm or totalizer pulse
    Transducers
    Construction
        Standard (FDT-41 thru 46): NEMA 6 (IP67), PVC, PEI, Nylon cord grip, PVC cable jacket
        Temperature: -40 to 60°C (-40 to 140°F)
        FDT-41 thru 46 with “-HT” option, FDT-47 and FDT-48: NEMA 6 (IP67), CPVC, PEI, Nylon cord grip, PVC cable jacket
        Temperature: -40 to 90°C (-40 to 194°F)
        FDT-47 “-HT”: NEMA 6 (IP67), PTFE, polyimide, Nickel-plated brass cord grip, PFA cable jacket
        Temperature: -40 to 176°C (-40 to 350°F)
    Frequency:
        FDT41 thru 46: 2 MHz
        FDT-47: 1 MHz
        FDT-48: 500 KHz
    Cables: RG59 coaxial, 75 Ω or twin axial, 78 Ω (optional armored conduit)
    Cable Length: Standard 6 m (20') 300 m (990') maximum in 3 m (10') incrementsSoftware Utilities Utilized to configure, calibrate and troubleshoot flowmeters. Connection via USB A/B cable; software is compatible with Windows 2000, Windows XP, Windows Vista® and Windows® 7
    RS485 Network All FDT-40 meters come equipped with RS485 drivers and utilize a MODBUS RTU command set (data can be returned in single-precision, doubleprecision, integer or floating point values). Up to 126 units products can be run on a single daisy-chain network and be individually queried for flow rate, positive flow accumulator, negative flow accumulator and signal strength. Flow accumulators can be cleared at discrete addresses or globally. The RS485 network is also compatible with direct to Excel, application.































暂无内容提供
暂无内容提供

联系方式

中国自动化网安全提示,在购买供应商商品时,请一定要先审核好企业信息是否真实,谨防网上诈骗。
公司名称:Omega工业测量(中国)
联系人: 王先生
联系电话:02161973664
手机号:
传 真:
电子信箱:kevin.wang@cn.omega.com
公司网站:http://cn.omega.com
公司地址:上海市宜山路1698号102单元
留言反馈
  • 评价:

  • 关于:

  • 联系人:

  • 联系电话:

  • 联系邮箱:

  • 需求意向:

  • 验证码:

    看不清楚?

  • 在线咨询
X
下载企业APP

成为企业会员免费生成APP!