  • Provides Relay Contact Closures at a Preset Temperature Input Level
  • Field Configurable Input Ranges for Type J, K, T, E, R, and S Thermocouples
  • Dynamic Deadband Prevents False Trips
  • Burnout Detection
  • Setpoints Programmable HI or LO
  • Selectable Failsafe/Latching Operation
  • Selectable 120/240 Vac Input Power

  • Model SMLA-TC of field configurable limit alarms with thermocouple input offers wide ranging inputs and flexiblesetpoint capability. It accepts six popular thermocoupletypes with inputs ranging from -270°C (454°F)to 1760°C (3200°F). The SMLA-TC provides twoindependent setpoint alarms (2 SPDT, 5A) andincludes 0.25% to 50% adjustable deadbands andselectable 120/240 Vac power.

    Model SMLA-TC is equipped with a dual functiondiagnostic LED. The green center LED indicates linepower and input signal status. Active line power isindicated by an illuminated LED. If the center LED isoff, check line power and the wiring connection. If theinput signal is above 100% full scale, the LED will flashat approximately 8 Hz. Below 0%, the flash rate isapproximately 4 Hz.

    Setpoints are top accessed multi-turn potentiometers.The field configurable SMLA-TC limit alarm setpointscan be configured for HI, LO, latching or fail-safetrip operation. Non-latching HI and LO setpointshave respective HI and LO deadbands. In a trippedcondition, the setpoint is exceeded and the appropriatered LED is lit. The trip will reset only when the processfalls below the HI deadband or rises above the LOdeadband. To reset a latched setpoint the signal mustbe in the safe region and the line power must beturned off for at least 5 seconds. For proper deadbandoperation, the HI setpoint must always be set above theLO setpoint.

    In failsafe operation, the relay is energized when theprocess is below a HI setpoint or above a LO setpoint(opposite for non-failsafe). In the failsafe mode, a powerfailure results in an alarm condition.

    The input must remain beyond the setpoint for100 milliseconds, uninterrupted, to qualify as a validtrip condition. Likewise, the input must fall outsidethe deadband and remain there for 100 millisecondsto return the alarm to an untripped condition. Thiseffectively results in a “dynamic deadband” - based ontime - in addition to the normal deadband.

    The factory default configuration for the SMLA-TCis as follows:
    Input: Type J, 0 to 360°C (32 to 680°F)
    Burn Out: Positive
    Output: Dual, SPDT
    Trip: A: HI, B: LO
    Latching: No
    Failsafe: No
    Deadband: A/B: 0.25%
    Power: 120 Vac

    When switching inductive loads, maximum relay life andtransient EMI suppression is achieved using externalprotection. Place all protection devices directly acrossthe load and minimize lead lengths. For AC inductiveloads, place a properly rated MOV across the load inparallel with a series RC snubber. Use a 0.01 to 0.1 mFpulse film capacitor (foil polypropylene recommended)of sufficient voltage, and a 47 Ω, ½ W carbon resistor.For DC inductive loads, place a diode across the load(PRV > DC supply, 1N4006 recommended) with (+) tocathode and (-) to anode (the RC snubber is an optionalenhancement).

    Specifications
    INPUT
    Ranges: Field configurable, see table
    Impedance: >1 MΩ
    Input Bias Current (Burnout Detect): <1.5 µA
    Overvoltage: ±10 V differential
    Common Mode (Input to Ground): 600 Vac or 1000 Vdc max
    LED Indications:
        Input Range (Green);
          >100% Input: 8 Hz flash
          <0% Input: 4 Hz flash
        Setpoint (Red): Tripped; Solid red (one LED for each setpoint)
    Limit Differential (Deadband): 0.25% to 50% of span
    Response Time:
        Dynamic Deadband: Relay status will change when propersetpoint/process condition exists uninterrupted
        for 100 msec
        Normal Mode (Analog Filtering): <250 msec (10 to 90%)
    Setpoint:
        Effectivity: Setpoints are adjustable over 100% of input span
        Repeatability (Constant Temp): ±0.2% for inputs > 0°C; ±0.3% for inputs < 0°C
    Stability:
        Line Voltage: ±0.01%/% max
        Temperature: ±0.05% of full scale/°C max
    Common Mode Rejection (DC to 60 Hz): 120 dB
    Isolation: 1000 Vdc between contacts, input and power
    ESD Susceptibility: Meets IEC 801-2, Level 2 (4 KV)
    Humidity:
        Operating: 15 to 95% RH @ 45°C (113°F)
        Soak: 90% RH for 24 hours @ 65°C (149°F)
    Temperature Range:
        Operating: 0 to 60°C (32 to 140°F)
        Storage: -15 to 70°C (5 to 158°F)
    Power: 120/240 Vac ± 10%, 50 to 60 Hz selectable, 2 W typical, 5 W max
    Relay Contacts: 1 SPDT (1 Form C) per setpoint
        Current Rating (Resistive):
          120 Vac: 5A
          240 Vac: 2A
          28 Vdc: 5A
        Material: Silver-Cadmium Oxide
        Electrical Life: 105 operations at rated load (external relay contact protection is required for use with inductive loads)
        Mechanical Life: 107 operations
    Latch Reset Time: 5 seconds (with power removed)
    Weight: 291 g (0.64 lb)































































暂无内容提供
暂无内容提供

联系方式

中国自动化网安全提示,在购买供应商商品时,请一定要先审核好企业信息是否真实,谨防网上诈骗。
公司名称:Omega工业测量(中国)
联系人: 王先生
联系电话:02161973664
手机号:
传 真:
电子信箱:kevin.wang@cn.omega.com
公司网站:http://cn.omega.com
公司地址:上海市宜山路1698号102单元
留言反馈
  • 评价:

  • 关于:

  • 联系人:

  • 联系电话:

  • 联系邮箱:

  • 需求意向:

  • 验证码:

    看不清楚?

  • 在线咨询
X
下载企业APP

成为企业会员免费生成APP!